Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
42
COI/AZ30E–EN Rev. D
…Calibrate
AutoCal Hardware
Selects the type of automatic calibration hardware to be
used.
Hardware Type
None
Disables automatic calibration functions.
None
Internal
Enabled if the (optional) built-in automatic calibration is
fitted.
Remote
Enabled if an external automatic calibration system is
used.
Note
. The solenoid valve outputs or relays can be used
to activate a remote automatic calibration system.
Test Gas Delay Time
The estimated delay in seconds it takes the test gas to
pass from a remote automatic calibration unit to the
probe.
Note
. For use with remote automatic calibration units
only.
[0 to 9999 s]
0
Test Gas Detection
The probe's internal automatic calibration system has
pressure switches to detect the presence of the test
gases automatically (when enabled). This option can
also be used with remote automatic calibration units with
suitable pressure switches.
Test Gas 1 Detection
Disabled if
AutoCal Hardware
/
Hardware Type
is set to
None
.
Disabled
Disabled
Enabled
If test gas is not present when a calibration is activated
an error message is created and calibration stops.
Test Gas 2 Detection
Disabled if
AutoCal Hardware
/
Hardware Type
is set to
None
.
Disabled
Disabled
Enabled
If test gas is not present when a calibration is activated
an error message is created and calibration stops.
Parameter
Comment / [Range]
Default
Table 6.7 Calibrate Menu (Continued)