Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
62
COI/AZ30E–EN Rev. D
…Input/Output
Current Output 1
Source
Selects the parameter to be retransmitted by the
current output.
% Oxygen
% Oxygen
Temperature
Cell mV
Type
Selects a linear or logarithmic output.
Note
. Displayed only when
Source
is set to
% Oxygen
.
Linear
Linear
Log. Two Decades
Log. Three Decades
Eng. Range Hi
Selects an engineering range high value, depending
on the option selected in
Source
.
Oxygen
[0.00 to 100.00 % O
2
*]
25.00 %O2
Temperature
[–200 to 1800 °C (–328 to 3272 °F)]
mV
[–100.0 to 400.0 mV]
Eng. Range Lo
Selects an engineering range low value depending on
the option selected in
Source
.
Note
. Not enabled if
Type
=
Log Two Decades
or
Log Three Decades
.
Oxygen
[0.00 to 100.00 % O
2
*]
0.25 %O2
Temperature
[–200 to 1800 °C (–328 to 3272 °F)]
mV
[–100.0 to 400.0 mV]
*The Endura AZ30 oxygen systems must be used only for measuring non-oxygen enriched gasses (not more than 21 %
oxygen [air]) at pressure no greater than 1.1 bar absolute (44 in. WG). Operating the system in oxygen-enriched
atmospheres will invalidate / compromise certification.
Parameter
Comment / [Range]
Default
Table 6.12 Input/Output Menu (Continued)