6.4.2 Calibrate
Used to calibrate the sensor, select test gas types, set oxygen
hold action, enable automatic calibration hardware (if fitted)
and set calibration diagnostics options.
Note
. Ensure reference air and test gas supplies are set up
before running a calibration – see Probe User Guide
OI/AZ30P–EN.
Parameter
Comment / [Range]
Default
Sensor Cal.
See Section
7.1
, page
69
for calibration routines.
Note
. Calibration options
1-Pt AutoCal
to
Manual Cal -
2-Pt
are displayed only when the probe reaches 690 °C
(1274 °F) and all error messages are cleared.
Restore Defaults
is displayed only when the probe
temperature is < 690 °C (1274 °F).
N / A
1-Pt AutoCal
Performs a single-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
2-Pt AutoCal
Performs a two-point automatic calibration.
Enabled only if calibration is currently inactive, Autocal is
fitted and temperature has stabilized.
Manual Cal – Offset
Performs a manual single-point offset calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Manual Cal – Factor
Performs a manual single-point factor calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Manual Cal – 2-Pt
Performs a manual two-point calibration.
Enabled only if calibration is currently inactive and when
temperature has stabilized.
Restore Defaults
Restores the factory calibration offset and factor values
entered in
Easy Setup
– see page 39.
Table 6.7 Calibrate Menu
2
Menu
Exit
Select
Calibrate