Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
10 Remote Computer Connection
98
COI/AZ30E–EN Rev. D
10.5 Cyclic Data
Cyclic data can be saved and output to a spreadsheet or as text. The update rate and data groups enabled
for output are selected at the
Communication / Cyclic Output
menu.
To output cyclic data:
1.
Launch a terminal application.
2.
Configure the terminal application to communicate with the COM port specified previously in the
Service Port Switch.
3.
To start the output routine, press 'P' on the PC keyboard (press 'S' to stop).
A header row is created and enabled data is output at the interval set.
4.
To output the data to a spreadsheet:
a.
use the capture capability to save a text (.txt) file
b.
open the .txt file into Excel
c.
at the Excel
Text Import Wizard
, select the file type as
Delimited
Data is labelled and formatted automatically for display and analysis.
10.6 Parameter Dump
The parameter dump option enables two types of information to be downloaded:
Configuration settings
Diagnostic Log and Signals View values
To run a
Configuration
settings dump:
1.
Launch a terminal application.
2.
Configure the terminal application to communicate with the COM port specified previously in the
Service Port Switch – see Section 10.4, page 97.
3.
Press 'C' on the PC keyboard to start the output routine.
4.
To output the data to a spreadsheet:
a.
use the capture capability to save a text (.txt) file
b.
open the .txt file into Excel
c.
at the Excel
Text Import Wizard
, select the file type as
Delimited
Data is labelled and formatted automatically for display and analysis.