Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
7 Calibration
72
COI/AZ30E–EN Rev. D
…Calibration Procedures
• New calibration
coefficients
discarded
•
Accuracy
check entry added
to Performance
Log
• Return to
Sensor
Cal.
• New calibration
coefficients saved
Save successful
•
Calibration
entry added to
Performance Log
• Return to
Sensor
Cal.
Save unsuccessful
•
Accuracy
check entry added to
Performance Log
• Display Failure
Calibration completed successfully
1-Pt
Aut
o
Cal
2-Pt
Aut
o
Cal
Manu
al
Cal
– Of
fset
Man
u
al
Ca
l – F
acto
r
Manua
l Cal – 2-Pt
• Return to
Sensor
Cal.
Calibration failed to complete
The reason for failure is displayed:
– stability criteria not met
– no response to the application of a
test gas detected
– calibration successful but
Save
of
new calibration factor unsuccessful
– newly calculated
calibration factor is outside the limits defined by
Factor Limit
and / or the newly calculated
calibration offset is outside the
Offset Limit
parameter – see page 46.
1-
Pt Au
toC
al
2-
Pt Au
toC
al
M
a
n
u
al
Cal –
Of
fset
M
a
nual
C
a
l
– Fac
tor
Man
u
al C
a
l – 2
-Pt
<Cal Type>
O2
1.00
%O2
Cal Offset
0.00 mV
Cal Factor
1.000
Reject
Accept
Reject
Accept
<Cal Type>
O2 20.95
%O2
Exit
Calibration Failed
Signal Not Stable
Exit
Signal Not Stable
No Response
NV Err
NV Error
Result Out of Bou
Result Out Of Bounds