Parameterization
D184B111U02 CoriolisMaster
FCM2000
83
Parameter
Value range / input type
Note
TOT2 Channel
TB Mass Flow
Only contained in the PROFIBUS PA software.
TB
Volume
Flow
TB
DiagExtMask
Bit-wise
selection
0x080001FE0FFF of
alarms
This can be used to select the bits of the error
register to be transmitted. For details, please see
our FCM2000 data link documentation
PROFIBUS PA.
Only contained in the PROFIBUS PA software.
CommSoftwareRev
2.11.0.12
Indicates the communication software version.
Only contained in the PROFIBUS PA software.
Dip
Switch
0
...
126
Technician*
Displays the switch position of the hardware
address switch for PROFIBUS PA after Enter is
pressed.
123456789A
Addr.
- -x - x- -x x x
20
Submenu
Function
test
In this menu you can test inputs and outputs as
well as the individual components of the
converter. Also included is a Simulation menu,
which provides the ability to set all measurement
variables in the converter to programmable
values.
Function
test
0.001
...
5000
Hz float
Pulse
Output
Technician*
Pulse Output
1 Hz
During normal operation the output of the pulses
is dependent on the flowrate selection (Mass or
Volume), the selected QmMax and the pulse
factor. In order to test the pulse output
independent of the converter settings, the self
check function permits the pulses per second
output on the pulse to be directly entered. The
value entered must be between 0.001 ... 5000
pulses/s. Pressing any key terminates the self
check process and the output is again a function
of the converter settings.
Not contained in the fieldbus software.
Function test
0 ... 26 mA
float
Iout
1
Technician*
Iout 1
10 mA
This function can be used to set a value for the
current output independent of the variable
assigned to the current output, for test purposes.
A current between 0 ... 26 mA can be simulated.
(In version A.00 the entry is made in percent)
Not contained in the fieldbus software.
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...