Parameterization
92 CoriolisMaster
FCM2000
D184B111U02
Calculation of accuracy
The accuracy of the concentration computation depends foremost on the quality of the matrix
data. As density and temperature measurement is the input for the calculation both accuracies
define the accuracy of the concentration measurement.
Example:
Density 0 % alcohol in water (20 °C [68 °F])
998.23 g/l
Density 100 % alcohol in water (20 °C [68 °F])
789.30 g/l
100 %
=
208.93 g/l
0.48 %
=
1 g/l
2.40 %
=
5 g/l
The chosen accuracy class of the density measurement effects the accuracy of the concen-
tration measurement directly.
Example: Input Matrix
10 °C (50 °F)
20 °C (68 °F)
30 °C (86 °F)
0 %
0 °BRIX
0.999 kg / l
0.982 kg / l
0.979 kg / l
10 %
10 °BRIX
1.010 kg / l
0.999 kg / l
0.991 kg / l
40 %
30 °BRIX
1.016 kg / l
1.009 kg / l
0.999 kg / l
80 %
60 °BRIX
1.101 kg / l
1.018 kg / l
1.011 kg / l
The unit of density and temperature corresponds to the chosen units in submenu units.
Pos: 41.13 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...