Installation
D184B111U02 CoriolisMaster
FCM2000
47
Pos: 33 /Überschriften/1.1/1-spaltig/D - F/Digitale Kommunikation @ 6\mod_1170934175875_3101.doc @ 66313 @ 22
4.7 Digital
Communication
Pos: 34 /==== Wechsel ein- auf zweispaltig ==== @ 0\mod_1130421847171_3101.doc @ 3828 @
Change from one to two columns
Pos: 35 /Technische Daten / Datenblatt/Durchfluss/FCM2000/Technische Daten/Messumformer/Digitale Kommunikation @ 5\mod_1167918524765_3101.doc @ 55897 @
The transmitter has the following options for digital communication:
4.7.1 HART
protocol
The system is registered with the HART Communication Foundation.
G00325
4 ... 20 mA
Rbmin = 250 Ohm
Bell 202
Modem Box
HART
Fig. 46:
Communication via HART protocol
HART protocol
Configuration
directly on the unit
Software DSV401 (+ HART-DTM)
Transmission
FSK modulation on voltage output
4 ... 20 mA acc. to Bell 202 standard
Max. signal
amplitude
1.2 mA
ss
Load
Current Output
min. 250
Ω
, max. = 560
Ω
(Dust-ignition proof: max. 300
Ω
)
Cable
Cable
AWG 24 twisted
Max. cable length
1,640.42 yd
Baud rate
1200 Baud
Display
Log. 1: 1200 Hz
Log. 0: 2200 Hz
For additional information, see the separate data link description.
System integration
In conjunction with the DTM (Device Type Manager) to the device,
the communication (configuration, parametrization) can occur with
the corresponding framework applications according to FDT 0.98 or
1.2 (DSV401 R2), respectively. Other desired tool/system inte-
grations (e.g., AMS or Siemens S7) are available upon request.
DSV401 communication tool for HART, free 90-day test version also
available upon request. DTMs are included in DSV401.
4.7.2 PROFIBUS PA protocol
The interface conforms to the Profile 3.0 (Standard PROFIBUS, EN
50170, DIN 19245 [PRO91]).
PROFIBUS PA ID no.:
0849 hex.
Alternative standard ID no.:
9700 or 9742 hex.
Configuration
directly on the unit
Software DSV401
(+ PROFIBUS PA-DTM)
Transmission signal
acc. to IEC 61158-2
Cable
shielded, twisted cable (acc. to
IEC 61158-2, types A or B are
preferred).
1 F
100
G00111
A
PROFIBUS DP
PROFIBUS PA
H2-Bus
PA+ PA- PA+ PA-
PA+ PA-
A = Segment coupler (incl. bus supply and termination)
Fig. 47:
Example for PROFIBUS PA interface connection
Bus topology
•
Tree and/or line structure
•
Bus termination: passive at both wire ends of the bus main wire
(RC-element R = 100
Ω
, C = 1 µF)
Voltage/current consumption
•
Medium current consumption: 14 mA.
•
In the event of an error, the integrated FDE function (=Fault
Disconnection Electronic) integrated in the device is ensures that
the current consumption can rise to a maximum of 26 mA.
•
The upper limit of the voltage is electronically limited.
•
The voltage on the bus line must lie in the range of 9 ... 32 V DC.
For additional information, see the separate data link description.
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...