Installation
D184B111U02 CoriolisMaster
FCM2000
37
Observe the following items when routing cables:
• The signal cable carries a voltage signal of only a few millivolts and therefore must be routed
the shortest distance possible. The maximum permissible signal cable length is 50 m
(164 ft.).
• Avoid routing the cable in the vicinity of electrical equipment or switching elements that can
create stray fields, switching pulses and induction. If this is not possible, run the signal cable
through a metal pipe and connect this to the station ground.
• All leads must be shielded and connected to station ground.
• Do not run the signal cable over junction boxes or terminal blocks.
• To shield against magnetic interspersion, the cable contains outer shielding that is attached
to the SE clamp.
G00413
1
Fig. 33
1
Make sure during installation that the cable is provided with a water trap (1). For vertical
installation, align the cable glands pointing downward.
Pos: 31.2 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...