Technical data
106 CoriolisMaster
FCM2000
D184B111U02
Pos: 53.4 /Überschriften/1.1/1-spaltig/M - O/Modell FCM2000-MS2 @ 11\mod_1172135843687_3101.doc @ 69589 @ 22
10.2 Model
FCM2000-MS2
Pos: 53.5 /Technische Daten / Datenblatt/Durchfluss/FCM2000/Technische Daten/Messaufnehmer/Modell FCM2000-MS2 @ 11\mod_1167918235609_3101.doc @ 55876 @ 5555
G00315
Fig. 67:
Flowmeter sensor for FCM2000-MS2
Meter sizes
"S" (DN 1.5); "T" (DN 3); "U" (DN 6)
Measuring ranges for flowrate
Nominal size
max. measuring range
[Q
max
] in [kg/min]
"S"
DN 1.5 (1/16“)
0 ... 100
"T"
DN 3 (1/10“)
0 ... 160
"U"
DN 6 (1/4“)
0 ... 475
Protection class: IP 65
Measurement value deviation for flowrate
± 0.4 % of measured value ± 0.02 % of Q
max
± 0.25 % of measured value ± 0.02 % of Q
max
± 0.15 % of measured value ± 0.01 % of Q
max
(deviation from measure value + zero error)
Reproducibility of flowrate
0.1 % of measured value for nom. deviation ± 0.15%
0.15 % of measured value for nom. deviation ± 0.25% and 0.4%
Measuring range density
0.5 ... 3.5 kg/dm
3
Measurement value deviation for density
Standard calibration ± 10 g/l
Temperature range 0 ... 100 °C (32 ... 212 °F
Expanded density calibration available upon request
Measurement value deviation for temperature
-50 ... 180 °C (-58 ... 356 °F) < 1 °C (33,8 °F)
Reference conditions
Calibration fluid
Water 25 °C (77 °F) (+ 5 K / - 5 K)
Pressure 0.5 ... 6 bar (7.3 ... 87.0 psi)
Ambient temperature
25 °C (77 °F) (+ 10 K / - 5 K)
Supply power
Line voltage per type plate U
N
± 1 %
Warm-up phase
30 minutes
Installation according to this specification
no visible gas phase
no external mechanical or hydraulic disturbances,
particularly cavitation
Output calibration
Pulse Output
Effect of the analog output on the measurement accuracy
Similar to pulse output ± 0.1% of measured value
Materials and additional technical data
Materials of flowmeter sensor
Fluid-wetted parts
1.4435 / 316L
Housing
1.4404
Fluid temperature
Standard:
-50 ... 180 °C (-58 ... 356 °F): DN 3 (1/10“), DN 6 (1/4“)
-50 ... 125 °C (-58 ... 257 °F): DN 1.5 (1/16“)
-50 ... 180 °C (-58 ... 356 °F): DN1.5 (1/16“) (optional)
For information about the design for operation in potentially explosive
area, refer to the corresponding chapter.
Ambient temperature
-20 ... 50 °C (-4 ... 122 °F)
For information about the design for operation in potentially explosive
area, refer to the corresponding chapter.
Process connections
G1/4" ISO 228-1
1/4" NPT ASME B1.201
Flange DIN/ASME for DN 6 (1/4“)
Threaded acc. to DIN 11851 for DN 6 (1/4“)
Tri-clamp DIN 32676 (ISO 2852) for DN 6 (1/4“)
The max. permissible operating pressure is specified in the
respective process connection, the fluid temperature, the screws and
the gasket material.
Pressure level
Flange PN 40, PN 100, Cl 150, Cl 600
Thread G 1/4", 1/4" NPT, PN 100 ... PN 410 (for each option)
Installation
For more detailed instructions regarding installation, refer to the
operating instructions.
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...