Parameterization
D184B111U02 CoriolisMaster
FCM2000
73
Parameter
Value range / input type
Note
Submenu
Totalizer
Submenu
Totalizer
Mass
This menu contains additional submenus for the
totalizers for the mass and volume flow integration
and one menu to simultaneously reset all the
totalizers.
All four totalizers (forward and reverse, mass and
volume flow totalizers) count to 10 million (in the
selected totalizer units). After a value of 10 million
is reached the corresponding overflow counter is
incremented by one and the totalizer value reset
to zero to continue totalizing the flow. In order to
indicate that an overflow has occurred a warning
is displayed. Up to 65535 overflows can be
registered. A value for each totalizer can be
individually set or reset (by entering a zero value)
in the appropriate menu. When a totalizer is set
(or reset), the relevant overflow counter
automatically resets to zero.
If the selection only „Forward“ was selected in the
menu Flow Direction in the submenu Operating
mode, then only the following totalizer menus are
displayed.
Counter
→
F
Display
123456789
kg
Input
Overflow
→
F
Display
0
Display of totalizer overflows for forward totalizer
Max. of 65535 overflows
Mass overflow counter
1 overflow = 10000000 Forward reset
Totalizer
←
R
Display
123456.78
kg
Input
Display of totalizer status for reverse direction;
displayed in the Forward/Reverse operating mode
Overflow
←
R
Display
Number of totalizer overflows
Max. of 65535 overflows
Mass overflow counter
1 overflow = 10000000;
displayed in the Forward/Reverse operating mode
Submenu
Totalizer
Volume
totalizer
Volume
Totalizer
→
F
Display
Display of totalizer status; forward totalizer
123456.78
kg
Input
Overflow
→
F
Display
0
Number of totalizer overflows;
1 overflow = 10000000
Totalizer
←
R
Display
123456.78
l
Input
Display of totalizer status for reverse direction;
displayed in the Forward/Reverse operating mode
Overflow
←
R
Display
0
Number of totalizer overflows
Max. of 65535 overflows
1 overflow = 10000000;
displayed in the Forward/Reverse operating mode
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...