Commissioning
D184B111U02 CoriolisMaster
FCM2000
57
Pos: 38.28 /Inbetriebnahme/Durchfluss/FCM2000/Hinweise für sicheren Betrieb - ATEX/Hinweise für sicheren Betrieb - ATEX @ 7\mod_1174468731953_3101.doc @ 73943 @
The safety specifications for the intrinsically safe circuits may be found in the EC-Type
Examination Certificate.
• Care must be exercised to assure that the protective cover over the supply power
connections is properly installed. For intrinsically safe circuits the connection area can be
opened.
• It is recommended that the appropriate cable connectors included with the shipment
be used for the output circuits: intrinsically safe blue; non-intrinsically safe black (not for
Version -40 °C [-40 °F]).
• The flowmeter sensor and the transmitter housing are to be connected to the potential
equalization. For intrinsically safe output circuits potential equalization is to be maintained
along the entire connection path.
• When the flowmeter sensor is insulated the max. insulation thickness allowed is
100 mm (4“).The transmitter housing may not be insulated.
• After the power to the flowmeter system is turned off, wait for a t > 2 min time period before
opening the transmitter housing.
• At start-up, EN50281-1-2 is to be considered for applications in areas with combustible dust.
• The operator must assure when protection ground (PE) is connected that no potential
difference can exist between the protection ground PE and the potential equalization PA,
even during a fault condition.
Special information for installations in Category 1:
• The interior of the meter pipe or meter sizes
≥
DN 50 (2") may be considered as Category 1
(Zone 0). The corrosion resistance of the materials must be considered.
Pos: 38.29 /Ex-technische Daten/Durchfluss/FCM2000/Hinweise zu Isolation @ 6\mod_1169211927484_3101.doc @ 61924 @ 3
5.3.4 Isolation: MC26.., MC27..
The pipeline and the flowmeter sensor insulation should be installed as shown.
The max. insulation thickness at the flowmeter sensor is 100 mm (4“).
G00331
1
Fig. 57
1 max. 100 mm (4“)
Pos: 38.30 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830 @
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...