Parameterization
D184B111U02 CoriolisMaster
FCM2000
65
Parameter
Value range / input type
Note
Unit name
Enter
Enter any unit name.
Baume
Min.
concentr. Enter
Enter
of min. valid concentration value.
0.00
Baume
Max. concentr.
Enter
Enter of max. valid concentration value.
100.00
Baume
Submenu
Matrix
1
Matrix
1
Matrix
2
Temperature
Enter all temperature values in chosen matrix.
Rules for valid values see 6.4.3.
Concentration
in unit
Enter all concentration values in unit in chosen
matrix. Rules for valid values see 6.4.3.
Concentration
in percent
Enter all concentration values in percent in
chosen matrix. Rules for valid values see 6.4.3.
Density
Enter the density values of the chosen matrix.
Entered values are marked with an "E".
Calculated values are marked with a "C". Rules
for valid values see 6.4.3.
Matrix calculation
Calculation of matrix based on above entered
values. Missing values are inter- or extrapolated.
Enter matrix
Store or discard of above values.
finish
Summary of Contents for CoriolisMaster FCM2000
Page 112: ...Appendix 112 CoriolisMaster FCM2000 D184B111U02...
Page 113: ...Appendix D184B111U02 CoriolisMaster FCM2000 113...
Page 114: ...Appendix 114 CoriolisMaster FCM2000 D184B111U02...
Page 115: ...Appendix D184B111U02 CoriolisMaster FCM2000 115...
Page 116: ...Appendix 116 CoriolisMaster FCM2000 D184B111U02...
Page 117: ...Appendix D184B111U02 CoriolisMaster FCM2000 117...
Page 118: ...Appendix 118 CoriolisMaster FCM2000 D184B111U02...
Page 119: ...Appendix D184B111U02 CoriolisMaster FCM2000 119...
Page 120: ...Appendix 120 CoriolisMaster FCM2000 D184B111U02...