5 Commissioning and First Operating Steps | 5.8 Working with Optional Accessories
ZEISS
7. Turn the knob counter-clockwise in order to
detach the specimen holder from the airlock
rod.
8. Retract the airlock rod.
5.8.1.2.2 Controlling the Valves via Airlock Macros
Procedure
1. In the
Airlock
panel, click
Resume Exchange
to start the pre-defined macro.
The macro automatically performs the following steps:
à
The gate valve is closed.
à
The column chamber valve is opened.
INFO:
The airlock macros can be modified
according to your individual requirements.
Contact your local ZEISS service representa-
tive for assistance.
2. Switch on the EHT. Refer to
.
9
You can now continue working with the microscope.
9
If you acquired an overview image with an airlock camera, you can now navigate the stage by
double-clicking in the overview image.
116