5 Commissioning and First Operating Steps | 5.6 Finding Appropriate Detector Settings
The following settings are recommended for the VPSE detector:
VPSE Bias
1 kV – 30 kV
7–10 mm
Adjustable from 0 to 100 %
Prerequisite
ü
The microscope is operating in VP vacuum mode
Procedure
1. In the SEM Controls panel, select the Detectors tab.
2. From the
Signal A
drop-down list, select
VPSE G4
.
3. Adjust the EHT, working distance (WD), and VPSE bias according to the suggestions in the
table in order to optimize the image.
5.6.6 Setting up the C2D Detector
image of radiolaria
Procedure
1. In the SEM Controls panel, select the
Detectors
tab.
2. From the
Signal A
drop-down list, select
C2D
.
3. Open the
Panel Configuration Bar
.
4. Double-click
C2D Control
to open the
C2D
Control
panel.
5. Select
C2D Gain = Low
.
6. Adjust the
C2D Bias
scroll bar until you reach saturation.
7. Click
C2D Auto Level
.
If you are unable to reach saturation or if the image quality is not good, repeat the proce-
dure in high gain.
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006