5 Commissioning and First Operating Steps | 5.8 Working with Optional Accessories
à
The column chamber valve is closed in order to separate gun head area and specimen
chamber.
à
The specimen stage is driven to the transfer position.
à
The airlock is evacuated.
à
The gate valve opens.
à
The illumination is turned on in the airlock
chamber.
5.8.1.1.2 Driving to Transfer Position without Airlock Macros
Prerequisite
ü
A specimen is loaded in the specimen chamber and is to be exchanged for another specimen.
Procedure
1. Switch off the EHT.
2. To open the
Stage Navigation Bar
, navigate to
View > Toolbars
and activate
Stage
Navigation Bar (for Widescreen users)
. Alternatively, you can access the
Stage Naviga-
tion Bar
via
Stage > Navigation
.
à
The
Stage Navigation Bar
is displayed.
3. From the
Panel Configuration Bar
, select
Airlock
.
4. Click
Close Column Chamber Valve
.
à
The column chamber valve is closed in order
to separate gun head area and specimen
chamber.
5. Select
Stage Points List
from the
Panel Configuration Bar
.
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006