What to do in the event of malfunctions
Version 9.2
G-30-1720-en
Page 127
Surgical field illumination
on microscope not
working.
Light guide not properly
inserted in microscope.
Insert the light guide as far
as it will go.
Failure of main and backup
lamps.
Replace both lamps.
Failure of stand electronics.
Use an OR lamp to illumi-
nate the surgical field.
Contact ZEISS Service.
Lamp module in stand has
no contact.
Insert the lamp module as
far as it will go.
Light source not activated
on the foot control panel.
Press light source on/off
button on the foot control
panel.
Suspension arm is in the
parking position.
Pull suspension arm
downward.
Insufficient surgical field
illumination.
Brightness level set too low.
Adjust brightness on stand
or foot control panel.
Lamp ageing may lead to
reduced light intensity.
Replace lamp.
Defective light guide
(illumination not uniform).
Replace the light guide.
Contact ZEISS Service.
-
Light guide not correctly
inserted.
Check that the light guide
is connected.
Filter selection knob is in an
intermediate position.
Turn knob until filter is
correctly in place.
Lamp is not properly seated
in the lamp socket.
Properly push lamp into
lamp socket.
Surgical field illumination
too bright.
Brightness level set too
high.
Adjust brightness on stand
or foot control panel.
Switch off light source on
stand. Use an OR lamp to
illuminate the surgical field.
Contact ZEISS Service.
Defective light control.
Switch to manual mode.
Red reflex too dark or
non-existent.
Red Reflex illumination not
activated.
Activate Red Reflex
illumination.
Contaminants on objective
lenses.
Clean objective lenses.
-
Malfunction
Possible cause
Remedy
Reference
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...