Version 9.2
Page 30
G-30-1720-en
Safety measures
OPMI Lumera
®
i on floor stand
Switch for the manual mode
3
Switch for the manual mode
In the event of failure of one of the major functions (XY movement, focus,
zoom, light control) and impairment of further functions, you can press
this switch to change to the manual mode.
The system reacts as follows:
–
All electrical components except the light source are disconnected
from the power supply.
–
The light source adjusts to a medium brightness level to permit
completion of the application.
–
The 5.7
″
control panel turns black.
To be able to continue surgery:
•
Manually set the magnification on the microscope’s zoom adjustment
knob (3.1) (e.g. with the help of a coin).
•
Manually position the stand for focusing and for moving the system in
the XY directions.
•
Use an external light source if the light source set to a medium bright-
ness level is not sufficient.
Press switch (3) again to return to normal operation. The switch unlocks,
and the system is restarted.
To ensure proper functioning of the manual mode when required,
perform status checks at regular intervals (at least every six months)
by activating the function (without a patient).
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...