Changing the halogen lamp
To change the lamp, proceed as follows:
•
Turn off the device with the power switch.
•
Press button (2) to eject lamp module (1) partially. Pull out the lamp
module.
•
Remove the defective halogen lamp from the spring-loaded mount.
•
Pull the ceramic base (8) from the contact pins of the halogen lamp.
•
Plug the ceramic base (8) onto the contact pins of the new halogen lamp.
•
Insert the new halogen lamp. Make sure you do not touch the lamp
bulb (7) or the interior of the reflector (6).
•
Press the halogen lamp into the spring-loaded mount (4)/(5).
•
Push the lamp module including the new halogen lamp back into the
device.
•
Turn on the device with the power switch.
3
“Instructions for use” sign
Observe the instructions for use or further applicable documents.
9
“CAUTION” warning label
Warning of a potential hazard.
•
Follow the instructions to change the lamp.
NOTICE
10 Risk of burns from hot lamp!
If you replace the lamp shortly after it has failed, the lamp will still be very
hot.
•
Allow the lamp to cool down before removing it.
•
Always use heat-resistant protective gloves when replacing the lamp!
NOTICE
Use the correct halogen lamp!
•
Only use the 12 V, 100 W halogen lamp specified in “Ordering data,
Consumables”.
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...