OPMI Lumera
®
i on floor stand
Operation
Version 9.2
G-30-1720-en
Page 97
•
Check lamp brightness
The lamp brightness can be controlled (see page 109).
The brightness adjustment has an effect on the surgical field illumina-
tion.
•
Check lamps
Primary lamp and backup lamp are both intact.
•
Check stand for secure positioning
At least two of the locking tabs of the casters have been pressed and
prevent the stand from starting to move by itself (see page 85).
•
Check user settings
The correct user profile has been set, and the settings have been
checked (see page 105).
•
Check unobstructed movement of the system
The system has been positioned in such a way that it can be moved
away from the patient at any time.
•
Check foot control panel
Stand and surgical microscope must perform the specified functions
(see page 112).
NOTICE
Risk of injury when tilting the surgical microscope!
If the surgical microscope is tilted heavily during an operation, the co-obser-
vation tube may swing downward unintentionally, or an accessory attached
to the surgical microscope may swing in unintentionally.
This can cause contact with and injury to the patient.
•
Always use extreme caution when adjusting the surgical microscope:
–
Hold the surgical microscope tight.
–
Loosen the friction knob only a little.
–
To the extent possible, tilt the surgical microscope only by a small
angle range.
–
Finally, tighten the friction knob firmly again.
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...