Operation
OPMI Lumera
®
i on floor stand
Version 9.2
Page 118
G-30-1720-en
Performing pairing
•
Press <System Settings> <System Options> <Pairing> in the main menu.
The “Pairing” menu is displayed.
•
Put the foot control panel in its vertical position in the immediate vicinity
of the system (max. distance of 1 m) and keep it in this position until pair-
ing is finished.
•
Start the pairing process by pressing the <Start pairing> button on the
5.7
″
control panel.
The following message is displayed: “Pairing of system and foot con-
trol panel is in progress. Press a button of the foot control panel until
the third LED from the top is flashing yellow AND keep the foot control
panel in a vertical position.”
Successful pairing
If pairing was successful, the “Radio link intensity” status indicator (1)
flashes green for approx. 1 s, and the following message appears:
–
“Pairing successfully completed. Bring the foot control panel in a hori-
zontal position (working position) and perform a function test. Set the
wheel of the foot control panel to the number indicated on the
suspension system.”
•
You can check whether pairing was successful by simultaneously pressing
any two buttons (2) on the foot control panel. The “Radio link intensity”
status indicator (1) then lights up.
Pairing failed
If pairing failed, the “Radio link intensity” status indicator (1) flashes red
for approx. 1 s, and one of the following messages appears:
–
“Pairing could not be successfully completed. No foot control panel
detected.”
–
“Pairing could not be successfully completed. Several foot control
panels detected.”
•
Repeat the pairing process as described above.
2
2
2
2
2
2
1
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...