Care and maintenance
OPMI Lumera
®
i on floor stand
Version 9.2
Page 132
G-30-1720-en
System maintenance
Electromagnetic compatibility
(EMC)
No regular inspections and maintenance are required in order to maintain
electromagnetic compatibility (EMC).
Maintenance intervals
Regular preventive maintenance is required in order to ensure safe and correct
operation of the device as well as a long service life.
The following table lists the relevant maintenance activities and shows at
which maximum intervals they must be performed.
The following sections describe the procedures to follow for performing main-
tenance tasks yourself.
Any other maintenance activities that are not described here in detail require
device-specific knowledge.
Please contact your local ZEISS Service or an authorized partner in a timely
manner to have these maintenance tasks carried out.
Interval
Maintenance work
Every 2 years
Visual inspection and function test of the following modules:
–
Stand with carrier arm
–
Illumination and power supply unit
–
Surgical microscope with binocular tube
–
Optical fiber
–
Mechanical brakes
–
Power cables
–
Connected accessories
–
Adjustment screw for limiting downward travel
–
OPMI coupling with securing screw
In accordance with country-
specific requirements
Safety check
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...