What to do in the event of malfunctions
OPMI Lumera
®
i on floor stand
Version 9.2
Page 130
G-30-1720-en
Zoom error. The error can
be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. jammed hardware
If the error impairs your
work, switch to the
manual mode.
Focus error. The error can
be acknowledged.
If the error impairs your
work, please switch to the
manual mode and contact
ZEISS Service.
–
e.g. jammed motor of
the focus system
If the error impairs your
work, switch to the
manual mode.
System boot error. Please
reboot the system. Please
make sure that no function
buttons are pressed during
the booting process.
–
e.g. a button was
pressed during system
start
Reboot the system. Make
sure not to press any
buttons during this
process.
-
System error. The error can
be acknowledged.
If the error recurs, please
contact ZEISS Service.
–
Unknown software or
hardware error.
Acknowledge the error
and contact ZEISS Service
if it occurs again.
-
Error message
Possible cause
Remedy
Reference
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...