What to do in the event of malfunctions
Version 9.2
G-30-1720-en
Page 125
Failure of a main function (XY, focus, zoom, light)
In the event that one of the major functions fails (XY movement, focus, zoom,
light control), leading to the impairment of further functions, you can switch
to the manual mode to complete a surgical procedure that has already been
started, for example.
•
Press button (1) to switch to the manual mode.
Working in the manual mode
In the manual mode, the light source is set to a medium intensity, and XY
coupling, focus drive, and zoom system are switched off.
•
Manually move the microscope in the X and Y directions for positioning.
•
Manually move the microscope up and down for focusing.
•
Use knob (2) to adjust the zoom value (magnification).
Fig. 36: Switching to the
manual mode
1
2
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...