Preparations for use
OPMI Lumera
®
i on floor stand
Version 9.2
Page 84
G-30-1720-en
Setting up the system
Limiting downward movement of the suspension arm
Proceed as follows to limit downward movement:
•
Loosen the locking lever (1) for the upward/downward movement.
•
Lower the surgical microscope until it can be focused on the surgical field
(depending on the focal length of the objective lens). At the same time,
ensure that there is sufficient safety distance to the surgical field.
•
Firmly retighten the locking lever (1) for the upward/downward movement.
•
Lower the surgical microscope to its lowest position again, and check the
safety distance.
The limit of downward travel is only effective in the lower part of the suspension
arm’s vertical movement range (starting from the arm’s horizontal position).
Fig. 29: Adjusting the
downward travel limit
CAUTION
Risk of injury caused by lowering of the surgical microscope!
If the surgical microscope has not been correctly balanced, it may lead to
downward movement and cause injury to the patient.
•
Adjust the limit of downward travel of the suspension arm in such a way
–
that the surgical microscope cannot come into contact with the
patient, even if it is inadvertently lowered.
–
that, during coarse focus with the suspension arm, there is sufficient
distance maintained from the patient, particularly if a fundus viewing
system and other accessories are mounted.
1
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...