OPMI Lumera
®
i on floor stand
Preparations for use
Version 9.2
G-30-1720-en
Page 71
Connection via video objective lenses VOL 442/802
Mounting the camera block
The following description is based on the video objective lens VOL 442.
The same process can be used for the video objective lens VOL 802.
•
Loosen the locking screw (4) on the video objective lens (5) by a few turns.
•
Insert the camera block (1) into the video objective lens (5) in the direction
of the arrow as far as it will go.
•
Firmly tighten the clamping screw (4).
Remove the aperture (3) before performing ophthalmic procedures.
•
For this purpose, remove the screw (2).
•
Remove the aperture (3).
The image brightness is increased.
The depth of field of the optical image is reduced.
Fig. 20: Mounting the
camera block
1
2
3
4
5
Содержание OPMI Lumera i on floor stand
Страница 1: ...ZEISS OPMI Lumera i on floor stand Instructions for use G 30 1720 en Version 9 2 2020 10 06 ...
Страница 38: ...Version 9 2 Page 38 G 30 1720 en Safety measures OPMI Lumera i on floor stand ...
Страница 41: ...OPMI Lumera i on floor stand Overview of devices Version 9 2 G 30 1720 en Page 41 Fig 4 System overview 3 1 2 ...
Страница 65: ...OPMI Lumera i on floor stand Preparations for use Version 9 2 G 30 1720 en Page 65 ...
Страница 99: ...OPMI Lumera i on floor stand Operation Version 9 2 G 30 1720 en Page 99 ...
Страница 122: ...Operation OPMI Lumera i on floor stand Version 9 2 Page 122 G 30 1720 en ...
Страница 143: ...OPMI Lumera i on floor stand Device data Version 9 2 G 30 1720 en Page 143 Fig 38 Dimensional drawing of OPMI Lumera i ...
Страница 166: ...Version 9 2 Page 166 G 30 1720 en Device data OPMI Lumera i on floor stand ...
Страница 176: ...Version 9 2 Page 176 G 30 1720 en OPMI Lumera i on floor stand ...
Страница 177: ...OPMI Lumera i on floor stand Version 9 2 G 30 1720 en Page 177 Blank page for your notes ...