Exchanging the lamp module of the xenon illumination system
• Turn off the suspension system at the power switch.
• Press button (2). The lamp module (1) is slightly ejected.
• Pull out the lamp module (1) as far as it will go.
• Slide the original transport case (3) over the lamp module. Bolt (4) must
engage in hole(5). This unlocks the stop.
• Remove the old lamp module and insert the new module.
• Turn on the suspension system at its power switch.
• Check the function of the xenon lamp and backup xenon lamp.
• Pack the old lamp module (6) in the transport packaging of the new lamp
module. Fill in the enclosed return card and send the old lamp module to
the nearest Zeiss service representative.
Please only use the original transport packaging (6), it is an explosion protec-
tion for eventually defective xenon lamps.
NOTE
Risk of burns caused by hot lamps!
If you change the lamp shortly after it has failed, the lamp will still be very
hot.
• Allow the lamp to cool down before changing it.
NOTE
Risk of injury!
The xenon lamp may fast burst due to the high inner pressure even if it is
cold. A hot lamp is therefore more dangerous due to the higher pressure and
the used materials.
In case of burst lamps, persons may be injured by flying glass splinters.
The lamp module shall only be changed by instructed staff.
• Only change the lamp module when it is cold.
• Avoid shocks at the lamp module.
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...