Version 7.0
G-30-1781-en
Page 213
S100 / OPMI pico
What to do in the event of malfunctions
Video image interference on
monitor or from video re-
corder.
Endoscope camera or video
equipment (monitor and
video recorder) on endo-
scope system defective.
Contact manufacturer of en-
doscope system.
–
Motion of surgical micro-
scope too stiff.
Friction adjustment screw on
stand tightened too firmly.
Loosen friction adjustment
screw on stand as required.
Problem
Possible cause
Remedy
Cross-
reference
Problem
Possible cause
Remedy
Cross
reference
No video image - only
color bars
Camera head not connected. Connect the camera head.
No image,
video indicator lamp is off,
Halogen lamp in light
source is lit
Plug of video power supply is
not properly plugged onto
the plate of the light source.
Contact service.
–
Defective light source.
Contact service.
–
Image with color tinge
No white balance
Perform white balance
Image without color
Y/C cable between CCU and
monitor not properly plugged
in.
Plug connector in properly.
Defective Y/C cable between
CCU and monitor.
Use different Y/C cable or BNC
cable.
No image, video indicator
lamp is lit
Monitor not switched on or
not correctly set.
Switch on monitor or set it to
correct channel.
Set monitor to default status, or
increase brightness and contrast.
Please refer
to monitor's
Instructions
for Use.
Monitor cannot display the
set resolution of the HD video
camera.
Use an HD compatible monitor
that can display the set resolu-
tion.
–
Connecting cable between
CCU and monitor or between
CCU and surgical microscope
not or not properly plugged
in.
Plug connector in properly.
Defective connecting cable
between video unit and mon-
itor.
Replace the cable.
–
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...