![Zeiss 000000-1080-349 Скачать руководство пользователя страница 212](http://html1.mh-extra.com/html/zeiss/000000-1080-349/000000-1080-349_instructions-for-use-manual_927636212.webp)
What to do in the event of malfunctions
S100 / OPMI pico
Version 7.0
Page 212
G-30-1781-en
Lamp in light source remains
dark and fan is running
Knob for brightness control
set to left-hand stop.
Turn knob clockwise.
Inclination switch deacti-
vated.
Move suspension arm of sur-
gical microscope down from
its standby position.
–
Inclination switch out of ad-
justment or defective
Contact service.
–
Lamp module has no contact. Insert lamp module to the
stop.
–
For halogen light source:
Defective halogen lamp.
Switch to backup lamp.
For halogen light source:
Ceramic base does not have
proper contact with halogen
lamp.
Plug ceramic base firmly onto
contacts of halogen lamp.
For xenon light sources:
Defective xenon lamp.
Switch to backup lamp.
For LED light source:
Defective LED light source.
Contact service.
–
Defective stand electronics.
Illuminate surgical area with
an OR luminaire.
Contact service.
–
Lamp goes constantly off and
on again during operation.
Ventilation openings are cov-
ered or dirty.
Ventilation openings must be
clear; clean them if necessary.
–
Defective fan.
Illuminate surgical area with
an OR luminaire. Contact
service.
–
Failure of suspension elec-
tronics.
Illuminate surgical area with
an OR luminaire. Contact
service.
–
Defective magnification
changer.
-
Contact service.
–
Filters in illumination beam
path are defective or cannot
be exchanged.
-
Contact service.
–
Problem
Possible cause
Remedy
Cross-
reference
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...