![Zeiss 000000-1080-349 Скачать руководство пользователя страница 247](http://html1.mh-extra.com/html/zeiss/000000-1080-349/000000-1080-349_instructions-for-use-manual_927636247.webp)
Version 7.0
G-30-1781-en
Page 247
S100 / OPMI pico
System Data
Electromagnetic immunity for ME equipment and ME systems
The S100 / OPMI pico is intended for operation in an electromagnetic environ-
ment as specified below. The customer or the user of the S100 / OPMI pico is
responsible for ensuring that the device is operated in such an environment.
Immunity
tests
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidelines
Electrostatic dis-
charge (ESD) as per
IEC 61000-4-2
±6 kV
contact discharge
±8 kV air discharge
±6 kV
contact discharge
±8 kV
air discharge
Floors should be made of wood or
concrete or be covered with ceramic
tiles. If the flooring contains synthetic
materials, the relative humidity must
be at least 30%.
Fast transient/burst
immunity as per
IEC 61000-4-4
±2 kV for
power supply lines
±1 kV for
input/output lines
±2 kV for
power supply lines
±1 kV for
input/output lines
The quality of the supply voltage
should be that of a typical business or
hospital environment.
Surges as per
IEC 61000-4-5
± 1 kV phase-to-neutral
voltage
± 2 kV phase/neutral to
ground voltage
± 1 kV phase-to-neutral
voltage
± 2 kV phase/neutral to
ground voltage
The quality of the supply voltage
should be that of a typical business or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
as per
IEC 61000-4-11
< 5% UT
(> 95% dip of UT) for
0.5 cycle
< 5% UT
(> 95% dip of UT) for
0.5 cycle
The quality of the supply voltage
should be that of a typical business or
hospital environment. If the user of
S100 / OPMI pico requires continued
function even in the event of inter-
ruptions in the power supply, we rec-
ommend to power the device S100 /
OPMI pico from an uninterruptible
power supply or a battery.
40% UT
(60% dip of UT) for 5 cy-
cles
40% UT
(60% dip of UT) for 5
cycles
70% UT
(30% dip of UT) for 25
cycles
70% UT
(30% dip of UT) for 25
cycles
< 5% UT
(95% dip of UT) for 5s
< 5% UT
(95% dip of UT) for 5s
Power frequency
(50/60Hz) mag-
netic field as per
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels typical of business
and hospital environments.
NOTE: UT is the AC supply voltage prior to application of the test level.
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...