Preparations for use
S100 / OPMI pico
Version 7.0
Page 104
G-30-1781-en
Adjusting the surgical microscope
Adjustment instructions
– In order to be able to quickly set/adjust the surgical microscope, it is ad-
visable to save the interocular distance as well as the refraction values for
the different users so they can be preset within the scope of the prepara-
tions.
– Do not wear multifocal or varifocal glasses during the adjustment process
or when working with the microscope later on.
Such glasses make a proper diopter adjustment ring setting impossible
and result in unsatisfactory imaging.
– Optimum image clarity in the eyepieces and at the optical port for docu-
mentation equipment (photo, video) can only be achieved through careful
focusing and precisely adjusting the binocular tube.
Adjusting the binocular tube
Adjusting the interocular
distance
• Move the microscope to a vertical position above a flat object, e.g. a sheet
of paper with writing.
• Adjust the distance of the eyepieces at the binocular tube to your interoc-
ular distance so that the two eyepiece images (object and edge of visual
field) merge into one.
Adjusting the eyepieces
The following procedure must be performed separately for each eyepiece in
the order described.
• Set the microscope to minimum magnification.
• Set the diopter adjustment ring at the eyepiece to zero position.
• Look through the eyepiece and move the microscope body to focus the
image.
• Set the microscope to maximum magnification and adjust the fine focus-
ing system until the image is sharply defined.
• Then reset the microscope to minimum magnification without changing
the working distance.
• Set the diopter adjustment ring at the eyepiece to its maximum value (plus
direction).
• Look through the eyepiece and slowly turn the diopter adjustment ring to
the minus direction until the image is sharply defined.
• Repeat the entire procedure for the second eyepiece.
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...