9-13
IM 704420-01E
Troubleshooting and Maintenance
9
9.6
Executing the Performance Test
CAUTION
• Do not apply a voltage exceeding the maximum input voltage to the input
terminal of each instrument. This may cause damage to the input section.
• Do not short the output terminals of instruments or apply external voltage to it.
This may cause damage to the internal circuitry.
Trigger Voltage (Slice Level) Accuracy Test
Items Required
The following items are required:
DC voltage generator
• Voltage accuracy: 1 mV or less
• Recommended instrument: Programmable DC voltage/current source 7651
(YOKOGAWA)
The procedure for testing the trigger voltage accuracy using the recommended
instruments is described below.
Connecting the Instrument
• Check that the power switch is turned OFF on all instruments before connecting
them.
• Connect the output of the DC voltage generator to the RF input connector of the
TA120E.
7651
PROGRAMAB LE DC SOURCE
YOKOGAWA
DATA
TRIG
TA120E
DC voltage generator (7651)
RF input connector
OUTPUT
Instrument settings
• TA120E
Measurement function: 3TCD
Trigger mode: MAN
• 7651
Output level: 4.000 V, 0 V, and –4.000 V
Note
If noise is present due to the influence from the outside environment, connect a 1-
µ
F
capacitor between the signal line and ground.