Outputting Signals, Initializing Setup Information, and Setting Key Lock
7
7.4
Initializing the Setup Information
There are two methods of initializing the setup information.
Keys
POWER
KEY
LOCK
MEAS
SET
AVE
/T
0
2
4
4
8
6
12
8
16
10
%
deg
%
20
180
us
ms
ns
MEAS
x1
3TCD
xN
3TCD
xN
BI-PHASE
SPEED
DELAY
GATE TIME
FUNC
GATE
LOCAL
DATA
SCALE
PRESET
PHASE
10
20
TRIG
CLOCK
PHASE
AUTO MAN
GP-IB
REMOTE
EQLZ
LEVEL
PLL
SHIFT
ARM
INH
DtoC
DVD
0.1s
0.5s
MAN
UTILITY
OPTION
10
5
σ
σ
Procedure
Note
•
Check that it is okay to initialize the settings before actually initializing the settings. You
cannot set the settings back after initialization. We recommend that you transmit the
setup information to a PC using a communication command and save the information
beforehand.
•
If you initialize the settings to factory default, the GP-IB address is also initialized.
Initializing the settings to factory default
1.
Check that the power switch is turned OFF.
2.
While pressing the MEAS key, turn ON the power switch. Hold the MEAS key
down for approximately 3 s. The settings are initialized.
Initializing the current setup information (settings excluding the GP-IB address
and the setup information that is stored in the internal memory)
1.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
2.
Press the > key. The settings are initialized. When the initialization is finished,
the characters [donE] are displayed for approximately a second, and the
measurement is resumed.