Procedure
Note
When calibration completes normally, a new calibration value is obtained. To apply the
calibration value obtained through this operation, you must change the factory default
calibration value. If you change the factory default calibration value, you cannot set it back.
Please confirm that it is okay to change the factory default calibration value beforehand. Note
that initialization (see section 7.4) will not set the calibration value back to factory default.
Carry out the operations below in the following cases:
• When the allowable range is exceeded in the performance test described in
section 9.6.
• When parts of the TA120E have been replaced.
Removing the signal cable
1.
Remove all cables that are connected to the TA120E (input and output cables)
excluding the power cable and GP-IB cable. Make sure that the TA120E is not
in the remote mode at this point.
For the procedure in switching from remote mode to local mode, see section 8.2.
Setting the TA120E in the maintenance mode
2.
Check that the power switch is turned OFF.
3.
While pressing the SCALE key, turn ON the power switch. Hold the SCALE key
down for approximately 3 seconds. After displaying [TA120E
→
704420
→
V-
x.xx(version display)
→
tESt
→
PASS, the characters [tEST] appear on the
display. The TA120E enters the start condition in the maintenance mode.
Let the TA120E warm up for at least 30 minutes in this condition.