3-7
IM 704420-01E
Measurement Preparation and Common Operations
3
3.4
Turning ON/OFF the Power Switch
Things to Check before Turning ON the Power
• Is the instrument properly installed?
→
Section 3.2, “Installing the Instrument”
• Is the power cord properly connected?
→
Section 3.3, “Connecting the Power Supply”
Location of the Power Switch and ON/OFF Operation
The power switch is located at the lower left corner of the front panel. The power switch
is a push button. Press once to turn it “ON” and press again to turn it “OFF.”
OFF
ON
Power Up Operation
When the power switch is turned ON, “[TA120E
→
704420]” appears on the 7-segment
LED display and the test program automatically starts. When the test program
completes normally, “[PASS]” is shown on the display and the TA120E is ready to make
measurements. The setup conditions are restored to the ones that existed immediately
before the power switch was turned OFF.
Note
If the TA120E does not operate as described above when the power switch is turned ON, turn
OFF the power switch and check the following points:
•
Is the power cord securely connected?
•
Is the correct voltage coming to the power outlet?
→
See section 3.3.
•
Is the circuit breaker ON?
→
See section 9.8.
•
You can initialize the settings of the TA120E. There are two methods of initialization. See
section 7.4.
•
Turn ON the power switch while pressing the MEAS key
→
the setup information is
initialized to the factory default condition.
•
Press the SHIFT+<(UTILITY) key and select [init]
→
the setup information excluding
the communication address and setup information stored to the memory is initialized.
If the instrument still fails to power up when the power switch is turned ON after checking
these points, it is probably a malfunction. Please contact your nearest YOKOGAWA dealer
as listed on the back cover of this manual for repairs.
To Make Accurate Measurements
Under the installation condition indicated in section 3.2, allow the instrument to warm up
for at least 30 minutes after the power switch is turned ON before starting the use of the
instrument.
Shutdown Operation
The setup information that exists immediately before the power switch is turned OFF is
stored. This holds true also when the power cord becomes unplugged. Note that the
measured results are not stored.
Note
The lithium battery that is used to store the setup information has a limited life span. When
the voltage level of the lithium battery drops below a given level, error code 909 appears on
the display when the power switch is turned ON. If the error code appears frequently, the
lithium battery must be replaced quickly. The user cannot replace the battery. Contact your
nearest YOKOGAWA dealer as listed on the back cover of this manual. For the life span of
the battery, see section 9.9.