9-10
IM 704420-01E
Performing calibration
4.
Press the LOCAL key. After the calibration items [dC
→
tV
→
PHASE
→
FunC] are
shown on the display, [PASS] or [FAiL] is displayed. This takes three to four
minutes.
If [PASS] is displayed, the calibration has been performed normally, and the
new calibration value has been obtained. To apply the new calibration value
(change the factory default calibration value), proceed to step 5. Otherwise,
proceed to step 14.
If [FAiL] is displayed, a problem has occurred. Proceed to step 6.
• Changing the factory default calibration value
5.
Press the SHIFT key. The characters [CoPY] appear on the display and the
factory default calibration value is changed. The new calibration value will take
effect the next time the power is turned ON. Proceed to step 14.
If the calibration fails to complete normally, [CoPY] is displayed and then [FAiL]
is displayed. In this case, servicing is required. Please contact your nearest
YOKOGAWA dealer as listed on the back cover of this manual for repairs.
Checking each of the calibration results
• Checking the DC voltage level
6.
Turn the rotary knob to select [dC].
7.
Press the > key. [PASS] or a hexadecimal value is displayed.
If [PASS] is displayed, the DC voltage level is normal.
If a hexadecimal value is displayed, a problem has occurred. Please note the
hexadecimal value that is displayed.
• Checking the time-voltage converter
8.
Turn the rotary knob to select [tV].
9.
Press the > key. [PASS] or a hexadecimal value is displayed.
If [PASS] is displayed, the time-voltage converter is operating normally.
If a hexadecimal value is displayed, a problem has occurred. Please note the
hexadecimal value that is displayed.
• Checking the programmable delay circuit
10.
Turn the rotary knob to select [PHASE].
11.
Press the > key. [PASS] or a hexadecimal value is displayed.
If [PASS] is displayed, the programmable delay circuit is operating normally.
If a hexadecimal value is displayed, a problem has occurred. Please note the
hexadecimal value that is displayed.
• Checking the signal multiplexer, the fractional pulse generator, and their
peripheral circuits
12.
Turn the rotary knob to select [FunC].
13.
Press the > key. [PASS] or a hexadecimal value is displayed.
If [PASS] is displayed, the signal multiplexer, fractional pulse generator, and
their peripheral circuits are operating normally.
If a hexadecimal value is displayed, a problem has occurred. Please note the
hexadecimal value that is displayed.
9.5 Performing Calibration (Changing the Factory Default Calibration Value)