Storing and Recalling Setup Information
6
Chapter 6 Storing and Recalling Setup Information
6.1
Storing the Setup Information
Keys
The preset number is displayed when setting
to store the setup information.
Lights when a preset number is shown on the display.
KEY
LOCK
MEAS
SET
AVE
/T
0
2
4
4
8
6
12
8
16
10
%
deg
%
20
180
us
ms
ns
MEAS
x1
3TCD
xN
3TCD
xN
BI-PHASE
SPEED
DELAY
GATE TIME
FUNC
GATE
LOCAL
DATA
SCALE
PRESET
PHASE
10
20
TRIG
CLOCK
PHASE
AUTO MAN
GP-IB
REMOTE
EQLZ
LEVEL
PLL
SHIFT
ARM
INH
DtoC
DVD
0.1s
0.5s
MAN
UTILITY
OPTION
10
5
σ
σ
Procedure
Selecting the preset number
1.
Press the SHIFT+SCALE(PRESET) key. The display shows the characters
[PrStx]. The preset number you selected previously is shown in [x]
2.
Turn the rotary knob to select the preset number for storing the setup
information.
Storing the setup information
3.
Press the > key. The characters [StorE] are displayed for approximately a
second, and the setup information is stored. When the setup information is
stored, measurement is resumed.
Explanation
Setup information that is stored
The following setup information is stored to the memory.
All settings such as the measurement function, gate, arming, inhibit, input, and
displayed unit excluding the GP-IB address.
Number of sets for storing presets
The number of sets that can be stored (preset number) is seven from 0 to 6.
Note
If you initialize all setup information of the TA120E to factory default settings (see section
7.4), the stored setup information is also initialized. The setup information of all seven sets is
reset to factory default settings.