Remote Control
[XR command] Queries the MAX range.
XR
Queries the MAX range.
Normal Adjustment
Command = XRm?<CRLF>
→
Return = XRmn<CRLF>
Parameters
m = 0: Internal pressure sensor
1: External pressure sensor
n = 0: 200 kPa range
1: 100 kPa range
2: 3500 kPa range
3: 16 MPa range
To set the user calibration date of the external pressure sensor, you
need to set Function to EXT P or EXT LEAK.
Yes
Yes
[YC command] Initializes measurement settings.
YC
Initializes measurement settings.
Normal Adjustment
Command = YCmn<CRLF>
→
Answer = YCm, OK<CRLF>
Parameters
m = 0: GROUP A item
1: GROUP B item
n = 1: Executes initialization
Yes
No
[ZC command] Initializes the zero calibration value.
ZC
Initializes the zero calibration value.
Normal Adjustment
Command = ZCm<CRLF>
→
Answer = ZC, OK<CRLF>
Parameters
m = 1: Executes initialization
Yes
No