5-15
IM CA700-01EN
1
2
3
4
5
6
7
Index
Remote Control
[SD command] Sets or queries the source value.
SD
Sets or queries the source value.
Normal Adjustment
Command = SDm<CRLF>
→
Answer = SDm<CRLF>
Command = SD?<CRLF>
→
Return = SDm<CRLF>
Parameters
m = Source value
DCmA m = 0.000 to 24.000
DCV m = 0.0000 to 5.5000
When the function is set to measurement, ERR13 is returned.
Yes
No
[SF command] Sets or queries the source function.
SF
Sets or queries the source function.
Normal Adjustment
Command = SFmn<CRLF>
→
Return = SFmn<CRLF>
Command = SFm?<CRLF>
→
Return = SFmn<CRLF>
Parameters
m = 1: GROUP B item
GROUP B item
n = 00: DCV
12: DCmA
16: mA Simulate
When the function is set to measurement, ERR13 is returned.
Yes
No
[SO command] Starts or stops or queries the source.
SO
Starts or stops or queries the source.
Normal Adjustment
Command = SOm<CRLF>
→
Answer = SOm<CRLF>
Command = SO?<CRLF>
→
Return = SOm<CRLF>
Parameters
m = 0: Source end (default value)
1: Source start
When the function is set to measurement, ERR13 is returned.
Yes
No
[SP command] Sets and queries % span.
SP
Sets and queries % span.
Normal Adjustment
Command = SPmn,o,p<CRLF>
→
Answer = SPmn,o,p<CRLF>
Command = SPm?<CRLF>
→
Return = SPmn,o,p<CRLF>
Parameters
m = 0: GROUP item A
1: GROUP item B
GROUP item A settings
n = 0: OFF (default value)
1: ON
o = Span lower limit
p = Span upper limit
-999999 to 000000 to 999999 (decimal point ignored)
GROUP item A settings
n = 0: OFF (default value)
1: ON
o = Span lower limit
p = Span upper limit
-999999 to 000000 to 999999 (decimal point ignored)
o and subsequent parameters are optional (when they are omitted, their
current settings remain unchanged).
Yes
No
5.4 Commands