3-4
IM CA700-01EN
3.3 Setting the Device Information and the
Calibration Conditions of the Pressure
Transmitter
Procedure
1.
In the calibration display, press
Edit (F1).
The calibration procedure setup screen appears. The setup screen is divided into three pages: Transmitter
Information, Measure, and Source.
Hold down ▲ or ▼ to change the screen.
Page
Setting the Device Information.
Entering the Model Number (Model No.)
2.
Use the ▲ or ▼ key to select
Model No.
, and press
ENTER
.
A character input screen appears.
Moves the
input position
Backspace
Confirms the input
Switches between “uppercase characters
and symbols” and “lowercase characters
and numbers”
Input position
Selected character ( (up, down, left, right) to move, ENTER to enter)
3.
Press (F1) or (F2) to move the input position, press ,
, ▲, or ▼ to select a character,
and press
ENTER
to enter the selected character.
4.
After entering the characters, press
Done (F5)
.
To discard the settings, press the
ESC
key.
Entering the Tag No. and Serial No.
5.
Use the ▲ or ▼ key to select
Tag No.
or
Serial No.
, and press
ENTER
.
6.
Like the model number, enter the tag number and serial number.