7-6
IM CA700-01EN
Adjusting the DC Current Measurement
2.
After step 1, use the ▲ or ▼ key to select
Current Measure
, and press
ENTER
.
The DC current measurement adjustment screen appears.
Input current
Switch the measurement range.
Adjusting the 20 mA Range
3.
Check that Input 20 mA is displayed, and apply 20 mA from the DC voltage/current generator.
4.
Press
Done (F5)
. The 20 mA range is adjusted.
Adjusting the 100 mA Range
5.
Press
Range (F1)
to set the measurement range to 100 mA.
Input 100mA appears.
6.
Check that Input 100mA is displayed, and apply 100 mA from the DC voltage/current generator.
7.
Press
Done (F5)
. The 100 mA range is adjusted.
Finishing the Adjustment
8.
Press
ESC
. The screen shown in step 1 appears.
7.2 Adjusting the Measurement Function