Source and Measure
2.1 Measuring Pressure
2-3
IM CA700-01EN
1
2
3
4
5
6
7
Index
Setting the Scaling
12.
Press the ▲ or ▼ key to select
Scaling
.
ON and OFF appear on the function menu.
13.
Press
ON (F1)
or
OFF (F2)
.
14.
If you select ON, simply press
ENTER
.
The scaling setting screen appears.
Set the lower span limit.
Set the upper span limit.
Set the lower scale limit.
Set the upper scale limit.
Set the decimal position of scale.
Set the unit.
Confirms the settings
Set the decimal position of span.
15.
Use the ▲ or ▼ key to select
Span Lower
, and press
ENTER
.
16.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set Span Upper, Scale Lower, and Scale Upper.
Span Upper must be set to a value greater than Span Lower.
Scale Upper must be set to a value greater than Scale Lower.
17.
Use the ▲ or ▼ key to select
Decimal Point
, and press
ENTER
.
The options of the span decimal point position are displayed.
18.
Use the ▲ or ▼ key to select a value, and press
ENTER
.
Note
The span decimal point position becomes the decimal point positions of Alarm and %Span.
19.
Repeat the operations similar to steps 16 through 18 to set the values of Scale Lower, Scale
Upper, and Decimal Point.
20.
Use the ▲ or ▼ key to select
Scale Unit
, and press
ENTER
.
A character input screen appears.
21.
Use the ▲, ▼,
, and keys to select a character, and then press
ENTER
. The character will
be entered.
After entering all the characters, press
DONE (F5)
.
22.
After setting the scaling parameters, press
Setting Done (F5
).
To discard all of the scaling settings, press the
ESC
key.
The system returns to the Display1 Setting display.
Note
The settings are not yet confirmed. To confirm the settings, press
Setting Done (F5)
again on the Display1
Setting display.