5-12
IM CA700-01EN
[MS command] Sets or queries the measurement/source mode.
MS
Sets or queries the measurement/source mode.
Normal Adjustment
Command = MSm<CRLF>
→
Answer = MSm<CRLF>
Command = MS?<CRLF>
→
Return = MSm<CRLF>
Parameters
m = 0: Measurement mode (default value)
1: Source Mode
Yes
No
[OD command] Requests to send measured data.
OD
Outputs the measured value.
Normal Adjustment
Command = ODm<CRLF>
→
Return = abbbbbbcsdddddddeee<CRLF>
m = 0: GROUP A item
1: GROUP B item
Parameters
<Header section (8 bytes)>
a = V: voltage/A: current/P: pressure
bbbbbb =
DC____:
DC
BAR___: bar
PAz___: Pa
ATMz__: atm
SCAL__:
when scaling is on
z: Gauge pressure G, absolute pressure A
c = N: normal/O: over-range/E: no data
<Data section (11 bytes)>
s = sign _: positive/-: negative (underscore represents a space)
ddddddd = measured value (7 digits)
eee = E+0/E+3/E-3/E+6/E-6
(for over-range, no data, or burnout:
dddddddeee=_999999.E+9)
When scaling is on, eee is “E+0.”
Output example
DCV: VDCN___0.000E+0
DCmA: ADCN___0.000E-3
Yes
Yes
[OE command] Outputs error information.
OE
Outputs error information.
Normal Adjustment
Command = OE<CRLF>
→
Return = ERRm<CRLF>
Outputs the most recent error.
After returning the error, the saved error number is overwritten with
“ERR00<CRLF>.”
“ERR00<CRLF>” is returned when there is no error.
Parameters
m =
Error code number
(See “Error Codes.”)
Yes
Yes
[PD command] Sets or queries the number of pressure display digits.
PD
Sets or queries the number of pressure display digits.
Normal Adjustment
Command = PDm<CRLF>
→
Answer = PDm<CRLF>
Command = PD?<CRLF>
→
Return = PDm<CRLF>
Parameters
m = 0: 6 digits (default value)
1: 5 digits
Yes
No