Source and Measure
Sweeping
The CA700 can vary the source value in the range defined by the lower and upper span limits. You can
vary the source value stepwise or linearly.
Step (Step)
The CA700 varies the source value in steps from the lower to the upper span limit.
Linear (Linear)
The CA700 varies the source value linearly from the lower to the upper span limit.
Span (Span Lower, Span Upper)
You can set the source range. The source range is defined by the 0% (Lower) and 100% (Upper)
values that you specify.
Number of Divisions (Number Of Points)
When sweeping in steps, set the number of divisions between the lower and upper span limits.
Source Time (Time Per Step)
When sweeping in steps, set the duration of each step. When sweeping linearly, set the duration for
the source value to change from 0% to 100%.
Sweeping in Steps
Duration of each step
Upper span limit
Lower span limit
Sweeping Linearly
Upper span limit
Lower span limit
Source time
Repeat Mode (Repeat Mode)
There are two repeat modes: Single Shot and Cont (Continuous).
Single Shot: The source changes from 0% to 100% to 0% and stops.
Cont: The source repeats the sweeping from 0% to 100% to 0%.
Saving Data
You can save the source value along with the pressure measurement data.
For details, see sections 4.1 and 2.1.
2.3 Generating DC Current and DC Voltage
2-17
IM CA700-01EN