Remote Control
[PU command] Sets or queries the pressure unit.
PU
Sets or queries the pressure unit.
Normal Adjustment
Command = PUmn<CRLF>
→
Answer = PUmn <CRLF>
Command = PUm?<CRLF>
→
Return = PUmn <CRLF>
Parameters
m = 0: DISPLAY1 item
1: DISPLAY2 item
n = 00: mbar
01: bar
02: Pa
03: hPa
04: kPa (default value)
05: MPa
06: atm
07: mmHg
08: inHg
09: gf/cm
2
10: kgf/cm
2
11: mmH
2
O@4°C
12: mmH
2
O@20°C
13: torr
14: psi
15: inH
2
O@4°C
16: inH
2
O@20°C
17: ftH
2
O@4°C
18: ftH
2
O@20°C
Yes
No
[PZ command] Executes zero calibration on the pressure value.
PZ
Executes zero calibration on the pressure value.
Normal Adjustment
Command = PZm<CRLF>
→
Answer = PZ,OK<CRLF>
Parameters
m = 0: Internal sensor
Yes
No
[REU command] Queries the user calibration date of the external pressure sensor.
REU
Queries the user calibration date of the external pressure sensor.
Normal Adjustment
Command = REU?<CRLF>
→
Return = REUyyyymmdd<CRLF>
Parameters
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
To set the user calibration date of the external pressure sensor, you
need to set Function to EXT P or EXT LEAK.
Yes
Yes
[RUD command] Queries the user calibration date of the CA700.
RUD
Queries the user calibration date of the CA700.
Normal Adjustment
Command = RUD?<CRLF>
→
Return = RUDyyyymmdd<CRLF>
Parameters
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
Yes
Yes
5.4 Commands