IM CA700-01EN
[DT command] Sets or queries the date/time.
DT
Sets or queries the date/time.
Normal Adjustment
Command = DTyyyymmddhhmmss<CRLF>
→
Answer = DTyyyymmddhhmmss<CRLF>
Command = DT?<CRLF>
→
Return = DTyyyy/mm/dd,hh:mm:ss<CRLF>
Parameters (default value)
yyyy: Gregorian (2013) 4 bytes, mm: month (01) 2 bytes, dd: day (01)
2 bytes
hh: hour (00) 2 bytes, mm: minute (00) 2 bytes, ss: second (00) 2
bytes
Yes
Yes
[ESC S command] Outputs the CA700 status byte.
ESC S
Outputs the CA700 status byte.
(”ESC” = ASCII 0x1B)
Normal Adjustment
Command = ESC S<CRLF>
→
Answer = m<CRLF>
Outputs the status byte as an integer.
Yes
Yes
[EUD command] Sets the user calibration date of the external pressure sensor.
EUD
Sets the user calibration date of the external pressure sensor.
Normal Adjustment
Command = EUDyyyymmdd<CRLF>
Parameters
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
To set the user calibration date of the external pressure sensor, you
need to set Function to EXT P or EXT LEAK.
Yes
Yes
[FCN command] Sets the user calibration date of the external pressure sensor.
FCN
Queries the connection state of the external pressure sensor.
Normal Adjustment
Command = FCN?<CRLF>
→
Answer = FCNm<CRLF>
Parameters
m = 0: Not connected
1: Connected
Yes
Yes
[FGD command] Queries the calibration date of the external sensor.
FGD
Queries the calibration date of the external sensor.
Normal Adjustment
Command = FGD?<CRLF>
→
Answer = FGDm,yyyymmdd<CRLF>
Parameters
m = 0: Not calibrated
1: Factory calibration date
2: User calibration date
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
To set the user calibration date of the external pressure sensor, you
need to set Function to EXT P or EXT LEAK.
Yes
Yes