5-16
IM CA700-01EN
[SY command] Switches between normal mode and user adjustment mode.
SY
Switches or queries the normal/adjustment mode.
Normal Adjustment
Command = SYm<CRLF>
→
Answer = SYm<CRLF>
Command = SY?<CRLF>
→
Return = SYm<CRLF>
Parameters
m = 0: Normal mode (default value)
1: User adjustment mode
Yes
Yes
[TD command ] Sets or queries the number of log points.
TD
Sets or queries the number of log points.
Normal Adjustment
Command = TDm<CRLF>
→
Answer = TDm<CRLF>
Command = TD?<CRLF>
→
Return = TDm<CRLF>
Parameters
m = 1 to 2000 (default value: 1)
Yes
No
[TM command] Sets or queries the data save method.
TM
Sets or queries the data save method.
Normal Adjustment
Command = TMm<CRLF>
→
Answer = TMm<CRLF>
Command = TM?<CRLF>
→
Return = TMm<CRLF>
Parameters
m = 0: SAVE (default value)
1: LOGGING
Yes
No
[TO command] Starts, stops, or queries the logging condition.
TO
Starts, stops, or queries the logging condition.
Normal Adjustment
Command = TOm<CRLF>
→
Answer = TOm<CRLF>
Command = TO?<CRLF>
→
Return = TOm<CRLF>
Parameters
m = 0: Logging stop (default value)
1: Logging start
Yes
No
[TR command] Sets or queries the logging interval.
TR
Sets or queries the logging interval.
Normal Adjustment
Command = TRm<CRLF>
→
Answer = TRm<CRLF>
Command = TR?<CRLF>
→
Return = TRm<CRLF>
Parameters
m = 0: 1 s (default value)
1: 2 s
2: 5 s
3: 10 s
4: 30 s
5: 60 s
Yes
No
[TS command] Executes manual saving.
TS
Sets the user calibration date of the CA700.
Normal Adjustment
Command = TS<CRLF>
→
Answer = TS,OK<CRLF>
Yes
No
[UD command] Sets the user calibration date of the CA700.
UD
Executes manual saving.
Normal Adjustment
Command = UDyyyymmdd<CRLF>
Parameters
yyyy: Year 4byte, mm: Month 2byte, dd: Day 2byte
Yes
Yes