Calibrating the Pressure T
ransmitter
3-5
IM CA700-01EN
1
2
3
4
5
6
7
Index
Setting Measurement Conditions
7.
Hold down the ▼ key.
The display switches to page 2/3 (measurement condition setting display).
When the measurement function is set to mA MEASURE,
mA LOOP, or V MEASURE
When the measurement function is set to P
(for details on P SWITCH, see section 3.6)
Turns averaging on and off
Turn scaling on and off and set scaling.
Set the measurement function.
Set the 0% value.
Set the 100% value.
Tolerance for pass/fail judgment
Confirms the settings
Confirms the settings
Turns averaging on and off
Turn scaling on and off and set scaling.
Set the measurement function.
Set the 0% value.
Set the 100% value.
Tolerance for pass/fail judgment
Set the unit.
When the measurement function is set to EXT P
(for details on EXT SWITCH, see section 3.6)
Confirms the settings
Turns averaging on and off
Turn scaling on and off and set scaling.
Set the measurement function.
Set the 0% value.
Set the 100% value.
Tolerance for pass/fail judgment
Set the unit.
Set the sensor type(fixed to 16MPa).
Set the measurement range.
Setting Measurement Functions
8.
Set the calibration target output.
Use the ▲ or ▼ key to select
Function
, and press
ENTER
.
The measurement function options are displayed.
9.
Use the ▲ or ▼ key to select the measurement function, and press
ENTER
.
If you set the measurement function to pressure (P or EXT P), the source function (SOURCE)
is set to mA SOURCE, mA SIMULATE, or V SOURCE.
If you set the measurement function to mA MEASURE, mA LOOP, or V MEASURE, the source
function (monitor only) is set to pressure (P or EXT P).
If the measurement function is set to pressure switch (P SWITCH or EXT SWITCH), you cannot
set the source function.
3.3 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter