5-14
IM CA700-01EN
[SC command] Sets and queries the scaling.
SC
Sets and queries the scaling.
Normal Adjustment
Command = SCmn,o,p,q,r,s,t,u<CRLF>
→
Answer = SCmn,o,p,q,r,s,t,u
<CRLF>
Command = SCm?<CRLF>
→
Return = SCmn,o,p,q,r,s,t,u <CRLF>
Parameter
m = 0: GROUP item A
1: GROUP item B
GROUP item A settings
n = 0: OFF (default value)
1: ON
o = Span lower limit
p = Span upper limit
-999999 to 000000 to 999999 (decimal point ignored)
q = Span decimal point position
0: 000000
1: 00000.0
2: 0000.00
3: 000.000
4: 00.0000
5: 0.00000
r = Scale lower limit
s = Scale upper limit
-999999 to 000000 to 999999 (decimal point ignored)
t = Scale decimal point position
0: 000000
1: 00000.0
2: 0000.00
3: 000.000
4: 00.0000
5: 0.00000
u = Scale unit
Up to 15 one-byte alphanumeric characters and symbols
that can be set on the main unit
GROUP item B settings
n = 0: OFF (default value)
1: ON
o = Span lower limit
p = Span upper limit
-99999 to 00000 to 99999 (decimal point ignored)
q = Span decimal point position (fixed to 0)
r = Scale lower limit
s = Scale upper limit
-99999 to 00000 to 99999 (decimal point ignored)
t = Scale decimal point position
0: 00000
1: 0000.0
2: 000.00
3: 00.000
4: 0.0000
u = Scale unit
Up to 15 one-byte alphanumeric characters and symbols
that can be set on the main unit
o and subsequent parameters are optional (when they are omitted, their
current settings remain unchanged)