3-8
IM AQ1210-01EN
Issuing a Work Completion Notice
WARNING
During measurement, light is transmitted from the light source ports. Do not disconnect the
connected optical fiber cables. Visual impairment may occur if the light enters the eye.
French
AVERTISSEMENT
Lorsque instrument génère de la lumière, la lumière est émise à travers les ports de source
lumineuse. Ne pas débrancher les câbles de fibre optique connectés. Des lésions oculaires
peuvent être causées si le faisceau lumineux pénètre l’œil.
3.
Press
REAL TIME
to start a real-time measurement.
During measurement, a mark appears at the top of the display to indicate that the laser light is on. Pressing
REAL TIME again stops the measurement.
4.
Tap the waveform display screen. A cursor appears at the tapped position.
If Notification is set to on in step 2, the word “CheckPoint” appears at the top area of the cursor.
A cursor appears at
the tapped position.
If work completion notification is turned on,
“CheckPoint” is displayed at the cursor.
Place the cursor to a
position that is a little
after the installation
point.
Work point
Set the
completion check
(OFF, ON).
Soft key menu display button
5.
Tap the soft key menu display button. The soft key menu is displayed.
6.
Tap the
Completion Check
soft key to select ON.
If the prior confirmation message is set to Only once or Check everytime in step 2 on page 3-7, when you
tap the Completion Check soft key, a message appears on the screen for you to confirm whether the cursor
has been moved to the work completion point.
When the measurement to the Check Point distance is complete, a completion message is displayed.
Note
You can perform measurements efficiently by moving the cursor to a position slightly after the installation
point (depending on how fast the optical fiber cable will be installed). This is because the instrument
notifies you of the work completion, which is your sign to measure the connection loss at that point (splice,
connection, etc.).
3.3 Cable Installation Completion Notification